Specification:
1. Diameter of curved finger: 12 mm
2. Bending finger length: 80mm (total length of three sections)
3. Diameter of baffle: 50mm
4. Baffle length: 100mm
5. Reference standard: IEC60065, IEC 60335-1, IEC 60529, IEC60884-1, IEC60950, IEC61032, IEC69745-1
The articulated test finger defined in IEC 61032 Figure 2 answers one question about an enclosure: can a finger reach a live part? This is that finger in its thrust version, the IEC 61032 50N Test Probe B, supplied as model HNT-PBT for the clauses that ask for the probe to be pressed home rather than merely offered to the opening.
The jointed sections are what make it a finger rather than a rod. Offered to a grille, a louvre or a recessed aperture, the three sections bend at their joints and follow the path a real finger would take, so the probe reaches behind the outer face instead of stopping at it. Behind the finger sit the baffle and the handle, which together keep the hand that is applying the force outside the boundary the test defines.
Labs keep it for the finger-touch and electric-shock checks that run through the appliance and electronic-equipment standards: household appliances, audio/video and IT equipment, socket accessories and fans. It is a bench instrument rather than a consumable, and one unit usually serves a whole product family.
| Model No. | HNT-PBT |
| Curved finger diameter | 12 mm |
| Bending finger length | 80 mm across three jointed sections |
| Baffle diameter | 50 mm |
| Baffle length | 100 mm |
| Thrust version | 50 N |
| Reference figure | IEC 61032 Figure 2 (Test Tool B) |
| Equivalent designations | IEC 950 Figure 2A; IEC 60884 Test Tool B |
| Standards referenced | IEC 60065; IEC 60335-1; IEC 60529; IEC 60884-1; IEC 60950; IEC 61032; IEC 69745-1 |
| Also cited against | UL 507; EN 60529 Figure 1; UL 1278 Figure 8.4 |
| Inspection purpose | Protection against finger touch; protection against electric shock |
| Specimen range | Accessible live parts or mechanical parts of the enclosure |
The finger body and the handle are a matte white plastic, with the tip left in polished metal and rounded off. Three sections make up the finger itself, joined with screws so a section can be re-angled by hand during insertion and will hold that angle until you change it again. A flanged collar separates the finger body from the handle, and the baffle disc forms the flat face that meets the enclosure.
Every part has a job. The metal tip is the working surface — it decides whether contact has been made, which is why it is left smooth and uncoated and why a burr or a flat spot is a real problem rather than a cosmetic one. The plastic body and handle keep the operator's hand outside the boundary the test defines and keep the probe light enough to hold steady while a force is being applied. The screw joints also let a laboratory strip, clean and reassemble the finger instead of retiring it the first time it picks up workshop grit.
The shaft carries a graduated scale for reading how far the finger has travelled into the opening, and the baffle edge is machined square so an applied force goes into the opening rather than being lost in the operator's grip. The result is a tool built so that two units of the same figure behave the same way on the bench.
Same body, different requirement. The finger you receive is the standard Figure 2 body — the same curved finger, the same bending length across three sections, the same baffle as the plain jointed finger B. What separates the IEC 61032 50N Test Probe B from that plain version is the force the test asks for, so the first thing to settle is whether your clause names one. If it does, order the thrust version and write it on the purchase order. If it does not, the plain finger covers the check.
Under a 50 N push an enclosure behaves differently. Flexible covers, soft gaskets, spring-loaded shutters and internal parts that sit a little proud can all give way under the applied force, and a probe that found nothing when it was lightly offered may reach a live part once it is pushed home. That is what the 50N force accessibility test for home appliances is checking: the enclosure under load as well as its geometry, and the test record should say which force was used so the number is not lost between reports.
Two practical notes. Because both versions share the same figure and the same dimensions, the force is the only thing that tells them apart on paper, which is why a jointed test finger manufacturer asks for your clause reference rather than a part description before quoting. And when a test is repeated after a design change, keep the same force and the same orientation as the original run, or the two results are not comparable.
On arrival. A probe is a measuring instrument, so the receiving check is a comparison with IEC 61032 Figure 2 rather than a glance at the finish. Straighten the three sections and check the bending length, tip diameter and baffle dimensions against the dimensional specification. Work every joint through its full arc by hand — a section should move when you ask it to and hold the angle you leave it in, because a joint that swings loose on its own will not carry the applied force. Inspect the metal tip under good light: round and smooth, with no flats, burrs or rolled edge, since that surface is what decides contact. Write the result on the receiving record and raise any mismatch with your supplier before the probe is used; a tool that has already appeared in a report is much harder to argue about than one still in its box.
Before type testing. Design and safety engineers can put this finger to work long before a certification project starts, and the thrust version is the one worth having on the bench at that stage. Offer it to a first-off moulding, a printed sample or a hand-built enclosure and the jointed sections show which louvres, recesses and apertures admit a finger at all; press it home at the applied force and the same openings are judged the way the type test will judge them, rather than the way a photograph suggests. A revision to a drawing at that point costs an afternoon; the same change once the tooling is cut costs a mould modification and a repeat run.
Keep the finger with the design team while the enclosure is still moving, and record what each run showed — which opening failed, from which direction, and under which force. Pass that to whoever is drawing the grille or the shutter, because aperture size is rarely the only variable: rib position, wall thickness and the space behind the face all change what the finger reaches. When the design does go forward to the type test, use the same probe, the same joint settings and the same orientations, so the design-stage sheet and the test report describe one set-up rather than two similar ones.
Choosing the right item starts with your requirement, not the catalogue. Our technical team works with buyers in Europe, North America and Asia to match a model to the clause or test condition you have to satisfy, and stays with you afterwards for commissioning, setup and troubleshooting. If an instrument already in service needs attention, we can supply replacement parts and advise on returning it to work. OEM and ODM requests go to the same team.
One finger covers finger access — the 2X numeral in an ingress protection sequence. Higher numerals need different tools: a 2.5 mm rigid rod for 3X and a 1.0 mm rigid wire for 4X. Most labs hold all three so a full sequence can run without stopping to look for equipment, and if you are building a set from scratch, tell the jointed test finger manufacturer which numerals you have to cover so the probes arrive matched to the same edition of the standard.
Compare it against the figure before it goes into service: curved finger diameter, bending length across the three sections, baffle diameter and baffle length should all line up with IEC 61032 Figure 2. Check that the joints move while still holding the angle you set, and that the metal tip is smooth. If anything looks off, raise it with your supplier straight away — a probe that enters a test record unchecked is a weak point in the report.
There is no service interval to track. The IEC 61032 50N Test Probe B stays in use until the metal tip is damaged, the jointed sections stop holding the angle you set, or a dimension check shows the probe has moved away from the figure. Treat a hard knock as the trigger for a check rather than something to carry on past until the next audit.
Yes, and that is often where it earns its cost. Design and safety engineers keep a finger on the bench through prototyping so that an opening which will fail the accessibility check is found on a mock-up rather than on finished tooling. Using the same tool at design stage and at type test also keeps the two sets of results comparable.
It shows how far the finger has entered the opening. When a test record has to describe insertion depth, or when two operators need to compare what they saw on the same grille, reading it off the shaft is faster than measuring afterwards. It is a reference on the instrument, not a substitute for checking the finger's dimensions against the figure.